1. Remove the Safe-Cone filter from the pipette.
2. Gently clean the surface of the pipette with a moistened cloth using a disinfectant liquid for example 70 %
ethanol, or mild detergent, and a soft, lint-free cloth. Pay special attention to cleaning the tip cone.
3. After cleaning, allow ethanol to dry, or remove detergent from the pipette thoroughly with a moistened cloth
using pure (Type 3) water or distilled water, then wipe it dry.
4. Insert the new Safe-Cone filter onto the pipette.
Cleaning & decontamination
Cleaning the lower part of the pipette
Set the cleaning and decontamination interval according to
the application. The recommendation is to clean every 3
months.
1. Check the cleaning or decontamination protocol and
chemical compatibility of the selected cleaning solution
for your pipette in the pipette user manual.
2. Wearing gloves, disassemble the pipette lower parts. As
a matter of routine, inspect the parts for wear and tear.
Replace the broken and worn parts with new ones or
send the pipette to a service provider.
3. Thoroughly clean the lower parts of the pipette with the
selected cleaning agent, and wipe the outer parts of the pipette handle. Don’t use tools to clean as that might
scratch the surface of the lower parts.
4. After cleaning, remove residual detergent, bleach, or
radioactivity cleaning agents from the pipette
thoroughly with pure (Type 3) water or distilled water.
Wipe the pipette dry or allow it to dry.
5. Re-grease the pipette. Use only the grease
recommended by the pipette manufacturer. Check the
performance of the pipette gravimetrically or with a
balance.
6. Pipette maintenance and performance test can be
outsourced from NOVA engineers.
Autoclaving
1. Remove the Safe-Cone filter if attached. Clean the
pipette with ethanol and allow it to dry.
2. Check the autoclaving protocol in the pipette user
manual to determine if the pipette is fully
autoclavable, or if only specific parts must be
disassembled and autoclaved.
3. Put the pipette or autoclavable pipette parts into
the sterilization bag, and place the bag into the
autoclave.
4. Sterilize at 121 °C and 1 bar overpressure for 20
minutes. (252 °F, 1 atm).
5. After autoclaving, cool down and dry the pipette
before reassembly. Drying can be done overnight at
room temperature or at 60 °C for 3 hours.
6. Insert the new Safe-Cone filter onto the pipette.
7. Reassemble the pipette, or check if the lower parts
have become loose during autoclaving, and tighten
them if necessary.
8. It is recommended to check the performance of the
pipette after every autoclaving. It is also
recommended to grease the piston or seal of the
pipette after every 10 autoclaves.